Table 5. Choice of objectives for Drosophila live cell imaging

Objective type Examplesa,b Specifications Applications
Dry 20X 0.75-NA, 40X 0.95-NA, 50X 0.5-NA LMPLFLN #1.5-coverslip-corrected, long working distance Deep imaging in thick tissues (e.g., embryos); micromanipulation and injection (upright microscope)
WI 60X 1.2-NA WI Coverslip correction collar, sensitive to coverslip angle, good working distance Deep imaging in thick tissues (e.g., embryos); multiphoton imaging
Water dipping 60X 0.9-NA, 100X 1.0-NA LUMPFL Upright microscopes, no coverslip, long working distance High-resolution imaging of live material in an upright configuration
Multi-immersion (glycerol) 20X 0.75-NA, 60X 1.3-NA 80% glycerol (better matched to RI of cytoplasm), coverslip correction collar, sensitive to coverslip angle Deep imaging in thick tissues (e.g., embryos); multiphoton imaging
Silicone oil 60X 1.3-NA silicone Requires silicone oil (better matched to RI of cytoplasm), coverslip correction collar, sensitive to coverslip angle Improved deep imaging in thick tissues (e.g., embryos)
Conventional oil 40X 1.3-NA to 100X 1.4-NA #1.5-coverslip-corrected, use a range of oils of varying RI to correct for spherical aberration Tissue-culture cells; imaging within maximum 30 μm of the coverslip; high-resolution imaging; fixed material
aExamples refer to Olympus objectives; in most cases, similar objectives are available from other manufacturers.
bAbbreviations: NA: numerical aperture; RI: refractive index; WI: water immersion.