Table

Table 1. Typical MuDPIT chromatography run setup

Step 1: 70-min gradient from 0% to 80% solvent B
10 min held at 80% solvent B
Steps 2-13: Each of these steps follows the sequence:
5 min of 100% solvent A
2 min of x% solvent Ca
3 min of 100% solvent A
10-min gradient from 0% to 10% solvent B
90-min gradient from 10% to 45% solvent B
Step 14: 5-min 100% solvent A wash
20-min 100% solvent C wash
5-min 100% solvent A wash
10-min gradient from 0% to 10% solvent B
90-min gradient from 10% to 45% solvent B
Step 15: 5-min 100% solvent A wash
20-min 100% solvent D wash
5-min 100% solvent A wash
10-min gradient from 0% to 10% solvent B
90-min gradient from 10% to 45% solvent B
aThe 2-minute solvent C percentages (x) in Steps 2-13 are as follows: 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, 90%, 90%, 100%, and 100%, respectively.