Table 1. Interpretation of PCR outcome

Template
Plasmid prepa Control yeastb Test clones Possible outcomes
x + + +
x +/−
x x + +/−
x +/−
Interpretation Bad mastermix/wrong settings/faulty amplifier Not enough template Lysed yeast inhibited PCR Uneven template load/digestion
Recommendation Double-check/repeat Add more template/improve lysis Add less template Adjust template load/re-PCR from obtained bands
Double bands may NOT be observed on the PCR done on the mixture of the plasmid prep with yeast, because smaller product from the empty vector may out-compete the bigger one from the plasmid.
aJG4-5-Raf1 is a suitable template; be sure to use a control library plasmid (with the insert) of the same type as is in the library.
bPCR from the empty vector yields a product of ~130 bp for JG4-5 (FP1 and FP2 primers); ~185 bp for YesTrp (YesTrp forward and reverse primers).